Autoclave (Vertical)
Vertical autoclave with Triple walled construction; chamber, basket, door lid, doorframe, bolts made of corrosion-resistant material and able to prevent stress cracking preferably made of high-grade stainless-steel sheet of SS-304 grade. Housing with SS legs
Pressure vessel should be Hydraulic tested at factory with minimum Hydrostatic Pressure: 2.5 kg/cm sq. (35 psi) ISI mark.
Fast safety lid lock with silicone gasket, it may be radial locking, automatic locking, single lever locking, fly nut assembly mechanism and with heat resistant/safety handle.
Manual water feed system with water level indicator, pressure gauge, steam release cock, spring loaded safety valve, water inlet and water valves.
Automatic Water Cut-off Device – To protect the heaters from running dry and to ensure that the machine is automatically switched off in case the desired water level falls below the prescribed level
Working temperature: 121°C, Maximum operating temperature: 134 °C (273 °F).
Working pressure: 15 PSI, Maximum operating pressure: 2.5 bar or 36 PSI Timer with Alarm System - To regulate the sterilization time of the media to be sterilized with a buzzer, Sterilization timer: 1–99 minutes.
A visual chamber gauge, which easily identifies pressure in the chamber must be accessible to the operator as a backup for reading pressure gauge when no electrical power is available.
Micro-processor temperature control system with sensor-with user changeable set temperature and timer option. The microprocessor controls the desired temperature (pressure automatically regulated) by cutting off the current to the heating element automatically & restart the mechanism as required.
The control panel to be mounted so that the components sensitive to steam and heat are protected
Equipment to work on 230 ± 10 volts single phase, 50 Hz, plug type adopted to local country scenario, Voltage regulator of appropriate rating to be included to cope with 160-260 V.
The line cord / Power cord supplied with the equipment shall be of acceptable durability, length, and current carrying capacity complying with Indian Standards. Remarks: The apparatus should confirm to national or international standards with latest amendments covering Markings, Safety requirements with recommendations of safe operations from any reputed firm with ISO 9001:2000 certification.
Accessories - Each Autoclave to be accompanied by an authorized list of accessories and spare parts. Accessories to be supplied include stainless steel basket (where 2 fit in autoclave directly plus two spare total 4), stainless steel wire basket (where 2 fit in autoclave directly plus two spare total 4), Chemical indicator tape for sterilization (2), Biological indicator (100), spare heating elements (two), fuses (10) and silicone gaskets (2).