Application: for Tissue Culture Laboratory and used for precise drying, heating, aging, burn-in, and hardening in various fields such as research, science, industry, and quality assurance.
Technical specification
Interior Dimensions (D x W x H) mm: 400 x 560 x 330 mm (±30mm)
Volume: 74 L
Housing: rear zinc-plated steel
Temperature:- Working Range: Min. 10°C to Max. 300°C
Setting range: Min. 20°C to Max. 300 °C
Setting accuracy: up to 99.9 °C: 0.1 / from 100 °C: 0.5
Sensor: 2 Pt100 sensors DIN Class A in 4-wire-circuit for mutual monitoring, taking over functions in case of an error
Ambient: Min. +5 °C to Max.+40 °C
Humidity rh: max. 80 %, non-condensing
Control technology:
Timer: Digital backwards counter with target time setting, adjustable from 1 minute to 99 days
Calibration: three freely selectable temperature values
Adjustable parameters: temperature (Celsius or Fahrenheit), fan speed, air flap position, programme time, time zones, summertime/wintertime-
Ventilation: Fan: forced air circulation by quiet air turbine, adjustable in 10 % steps
Fresh air: Admixture of pre-heated fresh air by electronically adjustable air flap
Vent: vent connection with restrictor flap
Safety
Temp. Control: mechanical temperature limiter TB, protection class 1
Over temperature monitor: Protect class 3.1
Alarm: visual and acoustic
Auto diagnostic system: for fault analysis
Auto Safety: additionally integrated over- and under temperature monitor
Standard equipment:
Works calibration: Calibration at +160°C
Door: fully insulated stainless steel door with 2-point locking (compression door lock)