Category: General Lab Equipment I

Test parameters: Minimum of 20 test parameters such as PT, APTT, Fibrinogen (FIB), thrombin time (TT), D-Dimer (DD), fibrin degradationproducts (FDP), Lupus Anticoagulant, Protein S, Protein C, Ant thrombin III, coagulation factors, Plasminogen Ac...



SPECIFICATION

  • Test parameters: Minimum of 20 test parameters such as PT, APTT, Fibrinogen (FIB), thrombin time (TT), D-Dimer (DD), fibrin degradationproducts (FDP), Lupus Anticoagulant, Protein S, Protein C, Ant thrombin III, coagulation factors, Plasminogen Activator inhibitor, von Will brand factors and other tests 
  • Throughput: minimum of 300 PT tests per hour and 300 PT and APTT tests per hour
  • Sample tube type: Use primary sample tube and secondary tubes
  • Test assay programmed: Pre-programmed and user definable assay programs 
  • Sample and reagent loading system: Capable of continuous sample and reagent loading during run
  • Measurement channels: Minimum of 20 measuring channels
  • Measuring Methods: clotting, chromogenic, immunoassay,
  • aggregation and other methods
  • Sample and reagent capacity: Minimum of 100 samples on board capacity positions with all STAT facility and a minimum of 40 reagent positions 
  • Sample and reagent volume: Has a system for sample and reagents level detection 
  • Dilution: Automatic dilution for sample and calibrators 
  • Access mode: Simultaneous random-access (samples, reaction tubes)
  • STAT Options: minimum of 5 priority positions
  • Sample running: Continuous sample & reagent loading, Auto redilution of samples, and able to add, delete, rerun tests during run. 
  • Reaction tubes/Cuvettes/test cups: Single reaction tubes and automatic continuous-access loading up to 1000 test cup in roll/pack/pcs
  • Quality control: Minimum of two level QC with capacity for plotting on Levy- Jennings graphs
  • Calibration: Automated calibration 
  • Software and Data Storage: Minimum of 10,000 sample test results 
  • Peripheral Ports: USB interface, support for external printers, hand-held barcode reader, Bi-directional LIS connectivity
  • Printer: Supplied with compatible external printer that uses cartilages available in the local market
  • Barcode reader: in-built barcode reader and Support for hand-held barcode reader identification of sample and reagents
  • Operating Environment: Operating Temperature (+15?C to +30?C), relative humidity <85% 
  • Utility Requirement: Electrical power supply: 220VAC ±10%, Suitable UPS with maintenance free batteries for minimum one-hour back-up
  • Standards & safety Requirements: Shall meet IEC-60601 general electrical safety and shall meet ISO 13485 standards (or equivalent BIS
  • Installation/Training Commissioning: providing installation, and commissioning of IQ, OQ and PQ for instrumentshe device at facilities, technical and end user training on site, Supplied with start-up reagent after installation