Hand Held Lcr Meter

Order Code: 19203326.18

Category: General Lab Equipment III

 FEATURE 20,000/2,000 Counts Dual Display. Test Frequency: 100/120Hz/1kHz/10kHz/100kHz. Auto LCR Mode for DUT Measuring. 0.2% Basic Accuracy. Measurement Parameters: L, C, R, (AC/DC), D, Q, ESR, . Parallel/Series Testing Mode. ...



SPECIFICATION

 FEATURE

  • 20,000/2,000 Counts Dual Display.
  • Test Frequency: 100/120Hz/1kHz/10kHz/100kHz.
  • Auto LCR Mode for DUT Measuring.
  • 0.2% Basic Accuracy.
  • Measurement Parameters: L, C, R, (AC/DC), D, Q, ESR, .
  • Parallel/Series Testing Mode.
  • Sorting Mode for Quality Control.
  • 2Wire or 5Wire Measurement Available.
  • Data Hold and Zero Mode Supported.
  • Max and Min.
  • Auto Range, Auto Backlit.
  • Low Battery Indication.
  • Auto Power Off.
  • Data Collection or DC Power Operation
  1.  SPECIFICATION
    1. Test Frequency: 100Hz/120Hz/1kHz/10kHz/100kHz Selectable.
    2. Full Scale: Main Display: 20,000/2,000 count selectable, Sub Display: 2,000 count.
  2. INDUCTANCE:
    1. Range: 20μH to 20kHz range depends on  the selected test frequency.
    2. Best Accuracy: ±(0.2% rdg + 2 digits).
    3. Resolution: 0.001μH to 0.001kH depends on the selected range.
  3. CAPACITANCE:
    1. Range: 20pF to 20mF range depends on the selected test frequency.
    2. Best Accuracy: ±(0.2% rdg + 2 digits).
    3. Resolution: 0.001pF to 0.001mF depends on the selected range.
  4. RESISTANCE:
    1. Range: 20 to 200M range depends on the selected test frequency.
    2. Best Accuracy: ±(0.2% rdg + 2 digits).
    3. Resolution: 0.001 to 0.01M depends on the selected range.
  5. DC RESISTANCE:
    1. Range: 200 to 200M in 7 ranges.
    2. Best Accuracy: ±(0.2% rdg + 2 digits).
    3. Resolution: 0.01 to 0.01M depends on the selected range.
  6. QUALITY FACTOR(Q)
    1. Range: 0.000 to 999.
    2. Accuracy: 2 x (main parameter accuracy).
    3. Resolution: 0.001
  7. DISSIPATION FACTOR (D)
    1. Range: 0.000 to 999.
    2. Accuracy: 2 x (main parameter accuracy).
    3. Resolution: 0.001.
  8. PHASE ANGLE ()
    1. Range: -90.0° to 90.0°.
    2. Accuracy: ±(0.2% rdg + 5 digits).
    3. Resolution: 0.1°.
    4. Measurement Circuit: Parallel or Series selectable.
    5. Auto LCR Mode: Automatically identifies and measures the DUT when the meter is switched on.
    6. Sorting Mode: ±0.1%, ±0.2%, ±0.25%, ±0.5%, ±1.0%, ±2.0%, ±5.0%, ±10.0%, ±20.0%, ±80%/-20% Selectable.
    7. Other Functions: Auto range, Auto back-light, Max, Min, Data hold, Zero, 46 segments analogue bar, Auto power off.
    8. Display: LCD mono display.
    9. Interface: USB.
    10. Power Source: AA Battery 1.5V x 4, DC 5V (throught AC adapter or USB cable).
    11. Dimensions Weight: 95(W) x 207(H) x 52(D) mm, Approx. 630g.
    12. Mandatory to attach a local office authorized letter which represented the oversea manufacturing to take care of after sales services, like calibration, local warranty repair and local certified engineer who had the certificate from manufacturer.

 ACCESSORIES User/Operational Manual x1
 INSTALLING AND COMMISIONING Installing and commissioning must be done within 1 month from the date of delivery at no additional cost from college community.
POWER REQUIREMENTS: 240 V, 50 Hz.

Enquiry Form