Order Code: 25268523C.1
Category: General Lab Equipment V
Parameter Details Name of the Equipment PLC Controlled Walk-in-Stability Test Chamber Control System Programmable Logic Control with Safety c...
|
Parameter |
Details |
|---|---|
|
Name of the Equipment |
PLC Controlled Walk-in-Stability Test Chamber |
|
Control System |
Programmable Logic Control with Safety controller, auto changeover of Standby systems & other event managements with 7” Color HMI with touch screen and Ethernet connectivity |
|
Memory |
200000 records |
|
Air Circulation |
Smart designed to achieve optimum energy utilization with integrated cooling, heating, and humidifying system. The air flow is designed to ensure uniformity of temperature under loaded conditions. |
|
Alarm |
Audio Visual Alarm if the temperature deviates from the preset temperature |
|
Lighting |
Interior illumination for working area |
|
Insulation |
80mm Thick CFC-free PUF Insulated modular cabinet |
|
Heating System |
U-shaped S.S. Nichrome wire air heaters |
|
Observation Window |
Full view see-through Glass Window |
|
Refrigeration System |
Cooling is achieved by a “AQUACHILL” chilled water refrigeration system |
|
Humidity System |
Non-condensing type steam injection system with water level arrangement and insulation to save heat energy |
|
Rack and Trays |
Specially designed adjustable rack and solid stainless-steel wire mesh trays ensure even distribution of temperature and humidity |
|
Power |
Works on 230/430 V AC single / three phase 50Hz |
|
Other |
MCB, Heavy duty latch with lock & key |
|
Safety Features |
1. Float switch to cut off mains in case of low boiler water level2. Safety thermostat to prevent overheating3. 180-sec compressor ON delay |
|
Specification |
Value |
|---|---|
|
Temperature Range |
20°C to 60°C |
|
Temperature Accuracy |
± 0.2°C |
|
Temperature Uniformity |
± 2°C |
|
Humidity Range |
40% RH to 95% RH |
|
Humidity Accuracy |
± 2% RH |
|
Humidity Uniformity |
± 3% RH |
|
ICH Conditions |
25°C/60%RH, 30°C/65%RH, 30°C/75%RH, 40°C/75%RH |