Refrigerated High Speed Centrifuge

Order Code: 25269410.13

Category: General Lab Equipment V

Refrigerated High Speed Centrifuge



SPECIFICATION

Refrigerated High Speed Centrifuge 

  1. A refrigerated microcentrifuge is an essential piece of laboratory equipment used to separate tiny volumes of temperature-sensitive samples by spinning them at high speeds.
  2. It employs centrifugal force to isolate components based on their density, while its cooling system protects fragile biological material from degrading due to the heat generated during rapid spinning
  3. Metal housing, chamber stainless steel

  4. Temperature set range from - 9 °C to + 40°C(Cooling from room temperature (20O C) to set temperature in just 9 minutes)

  5. Time set range 1 to 99 min, 1 min increments

  6. Standby cooling. - Induction motor, brushless drive

  7. Automatic rotor recognition and imbalance detection for maximum operational safety

  8. Acceleration/Deceleration time 12Sec/13 Sec, smooth, Rotors should acceleration up to 21,100 x g in 12 second’s time

  9. Rotor with aerosol-tight, transparent

  10. Rotor: The rotors should be resistance to mild biological acids

  11. Supplied with 24x1.5ml/2ml Rotor with click seal lid with 21,000 x g. The rotor should be tested and approved by Internationally Accredited Agency for Biocontainment.

  12. Supplied with 10 x 5 mL Rotor with Click Seal Biocontainment lid with 20,326 x g speed

  13. Maximum Speed above 14,600 RPM, Maximum RCF 21,000xg

  14. Max Noise Level: 50 dBA

  15. Programmable for all parameters (switchable between rpm and rcf) of a run

  16. Toggle between RPM and RCF.

  17. Input and recall of programs; at least 20 storage positions

  18. Large LED display for Time, Speed and Temperature

  19. Switchable, actual temperature, time left to finish run. - Imbalance switch-off. - Motor overheating protection.

  20. Chamber overheating protection

  21. Rotor recognition for appropriate over-speed protection

  22. Safety lock of lid during run and if the rotor is moving

  23. Possibility of mechanical opening of lid if there is current blackout

  24. Supply voltage: 230 ± 10 V, AC, 50/60 Hz.

  25. Voltage and plugs to be adapted to meet the country requirements.

  26. The line cord / Power cord supplied with the equipment shall be of acceptable durability, length, and current carrying capacity complying with Indian Standards. Power consumption.

  27. Approximately 1800 W.

  28. Conform to electrical safety standards IEC 60601–1, UL 61010–1, EN 61010–1. Excess-voltage category II.

  29. Protection class (in accordance with EN 60529). Designed not to interfere with circuit radio (in accordance with EN 55014).

Enquiry Form