Refrigerated High Speed Centrifuge
Metal housing, chamber stainless steel
Temperature set range from - 9 °C to + 40°C(Cooling from room temperature (20O C) to set temperature in just 9 minutes)
Time set range 1 to 99 min, 1 min increments
Standby cooling. - Induction motor, brushless drive
Automatic rotor recognition and imbalance detection for maximum operational safety
Acceleration/Deceleration time 12Sec/13 Sec, smooth, Rotors should acceleration up to 21,100 x g in 12 second’s time
Rotor with aerosol-tight, transparent
Rotor: The rotors should be resistance to mild biological acids
Supplied with 24x1.5ml/2ml Rotor with click seal lid with 21,000 x g. The rotor should be tested and approved by Internationally Accredited Agency for Biocontainment.
Supplied with 10 x 5 mL Rotor with Click Seal Biocontainment lid with 20,326 x g speed
Maximum Speed above 14,600 RPM, Maximum RCF 21,000xg
Max Noise Level: 50 dBA
Programmable for all parameters (switchable between rpm and rcf) of a run
Toggle between RPM and RCF.
Input and recall of programs; at least 20 storage positions
Large LED display for Time, Speed and Temperature
Switchable, actual temperature, time left to finish run. - Imbalance switch-off. - Motor overheating protection.
Chamber overheating protection
Rotor recognition for appropriate over-speed protection
Safety lock of lid during run and if the rotor is moving
Possibility of mechanical opening of lid if there is current blackout
Supply voltage: 230 ± 10 V, AC, 50/60 Hz.
Voltage and plugs to be adapted to meet the country requirements.
The line cord / Power cord supplied with the equipment shall be of acceptable durability, length, and current carrying capacity complying with Indian Standards. Power consumption.
Approximately 1800 W.
Conform to electrical safety standards IEC 60601–1, UL 61010–1, EN 61010–1. Excess-voltage category II.
Protection class (in accordance with EN 60529). Designed not to interfere with circuit radio (in accordance with EN 55014).