Application: for Tissue Culture Laboratory and used for maintaining precise and uniform temperature control in laboratory and industrial settings.
Technical specification
Dimensions (D x W x H) Interior: 400 x 560 x 480 mm (±50mm)
Volume: 108 L
Material: Stainless Steel
Housing: rear zinc-plated steel
Temperature: • Sensor: 1 Pt100 sensor DIN class A in 4-wire-circuit
Working Range: Min. 10°C to Max. 80°C
Setting range: Min. +20°C to Max. +80 °C
Setting accuracy: Approx. 0.1 °C
Ambient: Min. +5 °C to Max. +40 °C
Humidity: max. 80 %, non-condensing
Timer: Digital backwards counter with target time setting,
Adjustable from 1 minute to 99 days
Calibration: three freely selectable temperature values
Adjustable parameters: temperature (Celsius or Fahrenheit), fan speed, air flap position, programme time, time zones, summertime/wintertime
Fan: forced air circulation by quiet air turbine, adjustable in 10 % steps
Fresh air: Admixture of pre-heated fresh air by electronically adjustable air flap
Vent: vent connection with restrictor flap
Door: fully insulated stainless steel door with 2-point locking (compression door lock) and inner glass door
Internals Standard equipment: at least 2 stainless steel grid(s),
Electropolished
Programming: have software for reading out, managing and organizing the data logger via Ethernet interface (temporary trial version can be downloaded). USB stick with software available as accessory (on demand).
Power Supply: AC 220V-240V with 1400 W
Net weight: 76 kg (±6kg)
Standard Requirement:
Standard and Compliance: CE, EAC, UL, DIN 12880 or equivalent.
Supplier Requirement Experiences: at least 5 years.
Language: English language available on machine and manual.
Warranty: of Mechanical, Electrical parts and Services.
Include Installation and Training:
On-site installation performed by certified technicians.
User training covering equipment, full operation, maintenance, (minimum 1 full training session) at least 2 trainees.
Provision of training materials, including user manuals.